Please use this identifier to cite or link to this item: http://dr.iiserpune.ac.in:8080/xmlui/handle/123456789/6099
Title: Validity of point-mass model in off-resonance dynamic atomic force microscopy
Authors: RAJPUT, SHATRUHAN SINGH
DEOPA, SURYA PRATAP S.
AJITH, V. J.
KAMERKAR, SUKRUT C.
PATIL, SHIVPRASAD
Dept. of Biology
Dept. of Physics
Keywords: Viscoelasticity
Single-molecule force spectroscopy
Atomic force microscope (AFM)
Amplitude-modulation AFM
2021-JUL-WEEK3
TOC-JUL-2021
2021
Issue Date: Oct-2021
Publisher: IOP Publishing
Citation: Nanotechnology, 32(40), 405702.
Abstract: The quantitative measurement of viscoelasticity of nano-scale entities is an important goal of nanotechnology research and there is considerable progress with advent of dynamic atomic force microscopy. The hydrodynamics of cantilever, the force sensor in AFM measurements, plays a pivotal role in quantitative estimates of nano-scale viscoelasticity. The point-mass (PM) model, wherein the AFM cantilever is approximated as a point-mass with mass-less spring is widely used in dynamic AFM analysis and its validity, particularly in liquid environments, is debated. It is suggested that the cantilever must be treated as a continuous rectangular beam to obtain accurate estimates of nano-scale viscoelasticity of materials it is probing. Here, we derived equations, which relate stiffness and damping coefficient of the material under investigation to measured parameters, by approximating cantilever as a point-mass and also considering the full geometric details. These equations are derived for both tip-excited as well as base-excited cantilevers. We have performed off-resonance dynamic atomic force spectroscopy on a single protein molecule to investigate the validity of widely used PM model. We performed measurements with AFMs equipped with different cantilever excitation methods as well as detection schemes to measure cantilever response. The data was analyzed using both, continuous beam model and the PM model. We found that both models yield same results when the experiments are performed in truly off-resonance regime with small amplitudes and the cantilever stiffness is much higher than the interaction stiffness. Our findings suggest that a simple PM approximation based model is adequate to describe the dynamics, provided care is taken while performing experiments so that the approximations used in these models are valid.
URI: http://dr.iiserpune.ac.in:8080/xmlui/handle/123456789/6099
https://doi.org/10.1088/1361-6528/ac0cb1
ISSN: 0957-4484
1361-6528
Appears in Collections:JOURNAL ARTICLES

Files in This Item:
There are no files associated with this item.


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.